Troubleshooting
44
Troubleshooting
Problems you can handle yourself
If problems other than those described in the following tables arise, you must consult your nearest au-
thorized service.
Error
Behavior of the centrifuge
Possible cause(s) and measures to be taken
Displays remain dark
The motor stops.
The rotor stops without
braking.
The lid cannot be opened.
Mains failure or not connected
1. Check the mains connection.
2. If the mains connection is OK, call the nearest Service.
Displays fail briefly
The motor stops suddenly.
The rotor stops without
braking.
The display reads "br", see br.
Brief interruption of mains supply
1. Check whether the plug is plugged in properly.
2. Restart the centrifuge.
Lid cannot be opened
Pressing the "lid" key has no
effect.
Lid not correctly engaged or lid warped.
Press lid down in the middle of the front section.
Heat monitoring relays in the lid unlocking magnets have
been actuated
(repeated pressing of the "lid" key within a few seconds).
Press key again after a short pause (approx. 1 min).