Operation
24
Removing the rotor
To remove the rotor, you must follow the steps de-
scribed above in reverse order.
With the hermetic lid, you may in case of contamination
separate the rotor from the drive without opening the
lid! In this case you can open rotor upon removal from
the centrifuge using e.g. a safety work bench before
decontaminating it.
Danger of irreparable motor damage!
Never tilt the rotor. Always grab it in
the middle and pull out perpendicu-
larly.
1. Open the lid of the centrifuge.
2. Remove rotor cap (not necessary with the hermetic
lid).
3. Screw the cap nut open by turning it counterclock-
wise using the socket wrench delivered with the in-
strument. Remove the cap nut.
4. Grab the rotor in the middle and pull gently upwards
off the drive shaft. Be careful not to jam it.
Loading the rotor
As an accessory, you can optionally order the fixed-
angle rotor 7500 3325 with 24 places. Three sets of
adapters and a hermetic lid are available for this rotor
(see under "Accessories").
Improper or improperly combined ac-
cessories may cause severe damage
to the centrifuge!
Use only the parts and accessories
listed in the chapter "Accessories".