Troubleshooting
46
Error
Behavior of the centrifuge
Possible cause(s) and measures to be taken
"Lid" appears in the
display
Motor stops.
Rotor stops without braking.
Lid lock mechanism out of order
1. Press the lid shut. The centrifuge stops without braking.
2. If error persists, call nearest Service.
Display "OPEN“
appears although lid is
closed
Start impossible
Control lid lock mechanism. The lock must be engaged on
both sides.
If error persists, call nearest Service.
Flashing red point in
"time" panel
Displays remain dark
Not an error; instrument is on
standby
Press any key. With "start/stop" or "quick run", the centrifuge
starts running at once. With the "lid" key, the lid pops open,
and the display reads "OPEN".
In case you must call the Service
Should you require our Service, please tell us the order
no. and serial number of the instrument. You find the
pertinent information at the back of the instrument near
the socket for the mains plug.