The Biofuge haemo
7
The Biofuge haemo
The figure below shows the
Biofuge haemo
with the lid
opened. In this state, the standard display at the end of
a run and after opening the lid is "OPEN". After a cou-
ple of minutes the display changes to the standby
mode, and all display panels turn dark.
Safety systems
The
Biofuge haemo
is equipped with a number of
safety systems.
Rotor chamber
The rotor chamber is integrated into the steel case.
When the lid is closed, the rotor chamber is sealed
against the surroundings by a rubber ring with a spe-
cial profile that allows cooling air to pass through the
rotor chamber.
Warning if lid is manually opened during a run, or if
lid lock mechanism is out of order
If the lid is manually opened during a run, or if the lid
lock mechanism is out of order, a corresponding mes-
sage appears in the display ("Lid" and "OPEN",
respectively).
Lid lock mechanism
You can open the lid only when the power is turned on
and the rotor has practically come to a halt (< 80 rpm).
You can start the centrifuge only if the lid is properly
closed.
Emergency lid release
In order to permit you to remove samples even after a
power failure, the centrifuge is equipped with an emer-
gency lid release (see „Operation“).
ACHTUNG
ATTENTION
ATTENTION
ATENCIÓN
Rotor vorschriftsmäßig
auf Welle befestigen
Fixer le rotor sur
l'arbre conformément
aux préscriptions
Rotor must be fitted in
accordance with the
instruction manual
Fijar el rotor en el
arbol según las
prescripciones
Summary of Contents for Heraeus Biofuge haemo
Page 1: ...Instructions for use Biofuge haemo ...
Page 3: ...Biofuge haemo set set 1 2 0 0 0 1 0 speed time lid quick run start stop INSTRUMENTS ...
Page 10: ...For your safety 6 for your notes ...
Page 18: ...Accessories 14 for your notes ...
Page 20: ...Before use 16 for your notes ...
Page 41: ...Operation 37 for your notes ...
Page 51: ...Troubleshooting 47 ...