Before use
15
Before use
Where to install the centrifuge
The centrifuge must be operated in a place meeting the
following criteria:
•
A safety zone of 30 cm around the centrifuge must
be maintained. Hazardous materials must not be
kept within this zone during centrifugation.
•
The substructure must be stable and resonance-
free. A good support is provided by a plane labora-
tory bench or a large laboratory carriage with cast-
ers that may be locked.
•
To ensure sufficient air circulation, a minimum dis-
tance from the wall of 10 cm at the back and of
15 cm on each side must be kept.
•
The centrifuge must be protected from heat and
direct sunshine.
•
The location should be well ventilated.
Mains connection
Make sure that your mains voltage and frequency
match the specifications on the instrument.
Removing the transport protection
Turn the instrument on. The display panel shows for
about 6 s the routine internal software check sequence.
Open the lid by pressing the "open lid" key and remove
the transport protection for the rotor.
Check that the rotor moves freely by lightly turning it,
and make sure the rotor is tightly screwed on.