Operation
26
Aerosol-tight application
only with screw-on top 75003326 and
not with open container lids!
The following steps have to be carried out:
•
Lubricate the seals before inserting them (lubricant
order no. 75003500)
•
Insert the seal (C profile) in the groove at the side
of the body of the rotor.
•
Insert the O-ring into the inner groove on the
screw-on top.
Attention :
Please check that your sample containers are suitable
for the centrifugal application desired.
(16060 x g ; temperature in uncooled devices approx.
10 K above room temperature)
Please observe the permissible filling volumes!
Nominal volume:
Permissible volume:
2.0 ml
-
1.5 ml
1.5 ml
-
1.0 ml
others -
2
/
3
nominal volume
The sealing elements are to be checked regularly for
damage to the shape and surface!
Exchange faulty parts immediately (spare sealing rings
75003268)
The snap-on lid is not suitable for
aerosol-tight application!