Maintenance and care
41
Disinfection with bleaching lye
These agents contain highly aggres-
sive hypochlorites and must not be
used with aluminum rotors!
To protect the rotor 75003325 as far as possible you
must take the following precautions:
1. Avoid high temperatures!
The bleaching solution and the rotor should not be
warmer than ca. 25 °C.
2. Do not let the bleaching solution act longer than
absolutely necessary!
3. After disinfection, rinse the rotor thoroughly with
distilled water and allow to dry.
Decontamination
For general radioactive decontamination, use a solution
of equal parts of 70% ethanol, 10% SDS and water.
Follow this with ethanol rinses, then de-ionized water
rinses, and dry with a soft absorbent cloth.
Dispose of all washing solutions in appropri-
ate radioactive waste containers!