Maintenance and care
42
Autoclaving
Check whether autoclaving is permitted!
You may autoclave the rotor and the adapters at
121 °C.
Maximum permissible autoclaving cycle:
20 min at 121 °C.
For reasons of safety you may autoclave
the rotor 7500 3325 maximally 10 times!
The rotor must be cleaned and rinsed with distilled
water before being autoclaved. Remove the rotor lid,
the centrifuge tubes and the adapters. Place plastic
rotors on an even surface to avoid deformation.
Chemical additives to the steam are not
permitted.
Never exceed the maximum permis-
sible values for autoclaving tempera-
ture and autoclaving time.
Should the rotor show signs of wear,
you must stop using it!