Series 2600B System SourceMeter® Instrument Reference Manual
Section 2: General operation
2600BS-901-01 Rev. C / August 2016
2-97
Figure 47: LAN Status
1 LED indicates the LAN port is connected to a 100 Mbps network
2 LED indicates the LAN port is connected to a 10 Mbps network
When the LED is:
The network:
Off
is not connected
On
is connected
Blinking
is sending or receiving data
Using the LAN with remote operations
The following table lists the Series 2600B remote interface's available LAN protocols:
LAN protocols
Port number
Protocol
23
Telnet
1024
VXI-11
5025
Raw socket
5030
Dead socket termination port
You can only use one remote interface at a time. Although multiple ethernet connections to the
instrument can be opened, only one can be used to control the instrument at a time.
Raw socket:
Raw socket is a basic ethernet connection that communicates similarly to RS-232
without explicit message boundaries. The instrument will always terminate messages with a line feed,
but because binary data may include bytes that resemble line feed characters, it may be difficult to
distinguish between data and line feed characters.
VXI-11:
VXI-11 is similar to GPIB and supports message boundaries and service requests (SRQs). A
VXI-11 driver or VISA software is required. Test Script Builder (TSB) uses VISA and can be used with
the VXI-11 interface.
Telnet:
Telnet is similar to raw socket and is used when you need to interact directly with the
instrument, typically for debugging and troubleshooting. Telnet requires a separate telnet program.