Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-260
2600BS-901-01 Rev. C / August 2016
Also see
(on page 7-20)
Configuring and running sweeps
(on page 3-29)
(on page 7-260)
(on page 7-264)
(on page 7-266)
(on page 7-268)
(on page 7-269)
(on page 7-270)
(on page 3-32)
smuX.trigger.measure.action
This attribute controls measurement actions during a sweep.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Not saved
0 (smuX.DISABLE)
Usage
action
= smu
X
.trigger.measure.action
smu
X
.trigger.measure.action =
action
action
The sweep measurement action; set to one of the following values:
0
or
smu
X
.DISABLE
: Do not make measurements during the sweep
1
or
smu
X
.ENABLE
: Make measurements during the sweep
2
or
smu
X
.ASYNC
: Make measurements during the sweep, but asynchronously
with the source area of the trigger model
X
Source-measure unit (SMU) channel (for example,
smua.trigger.measure.action
applies to SMU channel A)
Details
With this attribute enabled (setting
action
to
smu
X
.ENABLE
or
smu
X
.ASYNC
), configure the
measurement with one of the
smu
X
.trigger.measure.
Y
()
functions.
If this attribute is set to
smu
X
.ASYNC
:
•
Asynchronous sweep measurements can only be used with measure autoranging turned off. To turn
measure autoranging off for all measurements during the sweep, set the
smu
X
.measure.autorange
Y
attribute to
smu
X
.AUTORANGE_OFF
.
•
Autozero must also be turned off. To turn off autozero, set the
smu
X
.measure.autozero
attribute to
smu
X
.AUTOZERO_OFF
or
smu
X
.AUTOZERO_ONCE
.
If either of the above items is incorrectly configured, the
smu
X
.trigger.initiate()
function
generates an error.
Also see
(on page 7-253)
(on page 7-264)
(on page 3-32)