Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-320
2600BS-901-01 Rev. C / August 2016
Usage
operationRegister
=
status.operation.instrument.tsplink.trigger_overrun.condition
operationRegister
= status.operation.instrument.tsplink.trigger_overrun.enable
operationRegister
= status.operation.instrument.tsplink.trigger_overrun.event
operationRegister
= status.operation.instrument.tsplink.trigger_overrun.ntr
operationRegister
= status.operation.instrument.tsplink.trigger_overrun.ptr
status.operation.instrument.tsplink.trigger_overrun.enable =
operationRegister
status.operation.instrument.tsplink.trigger_overrun.ntr =
operationRegister
status.operation.instrument.tsplink.trigger_overrun.ptr =
operationRegister
operationRegister
The status of the operation status TSP-link overrun register; a zero (0) indicates no
bits set (also send 0 to clear all bits); other values indicate various bit settings
Details
These attributes are used to read or write to the operation status TSP-link overrun registers. Reading
a status register returns a value. The binary equivalent of the returned value indicates which register
bits are set. The least significant bit of the binary number is bit B0, and the most significant bit is bit
B15. For example, if a value of
10
is read as the value of the condition register, the binary equivalent
is 0000 0000 0000 1010. This value indicates that bit B1 and bit B3 are set.
B15 B14 B13 B12 B11 B10 B9
B8
B7
B6
B5
B4
B3
B2
B1
B0
**
>
>
>
>
>
>
>
>
>
>
>
>
>
>
*
0
0
0
0
0
0
0
0
0
0
0
0
1
0
1
0
* Least significant bit
** Most significant bit
A set bit indicates that the specified line generated an action overrun when triggered to generate an
output trigger.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Enable and transition registers
(on page E-19). The individual bits of this
register are defined in the following table.
Bit
Value
Description
B0
Not used
Not applicable
B1
status.operation.instrument.tsplink.trigger_overrun.LINE1
Bit B1 decimal
value: 2
B2
status.operation.instrument.tsplink.trigger_overrun.LINE2
Bit B2 decimal
value: 4
B3
status.operation.instrument.tsplink.trigger_overrun.LINE3
Bit B3 decimal
value: 8
B4-B15
Not used
Not applicable