Series 2600B System SourceMeter® Instrument Reference Manual
Section 2: General operation
2600BS-901-01 Rev. C / August 2016
2-85
Range programming
Range commands
The following tables summarize commands necessary to control measure and source ranges. See
the
(on page 7-1) for more details about these commands.
Measure range commands*
Command**
Description
smu
X
.measure.autorangei = smu
X
.AUTORANGE_ON
Enable current measure autorange.
smu
X
.measure.autorangei = smu
X
.AUTORANGE_OFF
Disable current measure autorange.
smu
X
.measure.autorangev = smu
X
.AUTORANGE_ON
Enable voltage measure autorange.
smu
X
.measure.autorangev = smu
X
.AUTORANGE_OFF
Disable voltage measure autorange.
smu
X
.measure.lowrangei = lowrange
Set lowest I measure range for
autorange.
smu
X
.measure.lowrangev = lowrange
Set lowest V measure range for
autorange.
smu
X
.measure.rangei = rangeval
Select manual current measure range.
smu
X
.measure.rangev = rangeval
Select manual voltage measure range.
* See
**
smu
X
: For Models 2601B, 2611B, and 2635B, this value is
smua
(SMU Channel A); for Models 2602B, 2604B, 2612B,
2614B, 2634B, and 2636B, this value can be
smua
(for SMU Channel A) or
smub
(for SMU Channel B).
Source range and limit commands*
Command**
Description
smu
X
.source.autorangei = smu
X
.AUTORANGE_ON
Enable current source autorange.
smu
X
.source.autorangei = smu
X
.AUTORANGE_OFF
Disable current source autorange.
smu
X
.source.autorangev = smu
X
.AUTORANGE_ON
Enable voltage source autorange.
smu
X
.source.autorangev = smu
X
.AUTORANGE_OFF
Disable voltage source autorange.
smu
X
.source.limiti = level
Set voltage source current limit
(compliance).
smu
X
.source.limitv = level
Set current source voltage limit
(compliance).
smu
X
.source.limitp = level
Set source power limit (compliance).
smu
X
.source.lowrangei = lowrange
Set lowest I source range for autorange.
smu
X
.source.lowrangev = lowrange
Set lowest V source range for autorange.
smu
X
.source.rangei = rangeval
Select manual current source range.
smu
X
.source.rangev = rangeval
Select manual voltage source range.
* See
**
smu
X
: For Models 2601B, 2611B, and 2635B, this value is
smua
(SMU Channel A); for Models 2602B, 2604B, 2612B,
2614B, 2634B, and 2636B, this value can be
smua
(for SMU Channel A) or
smub
(for SMU Channel B).