Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. C / August 2016
7-369
trigger.blender[N].stimulus[M]
This attribute specifies the events that trigger the blender.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Trigger blender N reset
Configuration script trigger.EVENT_NONE
Usage
eventID
= trigger.blender[
N
].stimulus[
M
]
trigger.blender[
N
].stimulus[
M
] =
eventID
eventID
The event that triggers the blender action; see
Details
N
An integer representing the trigger event blender (1 to 6)
M
An integer representing the stimulus index (1 to 4)
Details
There are four stimulus inputs that can each select a different event. The
eventID
parameter can be
the event ID of any trigger event.
Use zero to disable the blender input.