smuX.trigger.endsweep.action
This attribute sets the action of the source at the end of a sweep.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Not saved
0 (smuX.SOURCE_IDLE)
Usage
action
= smu
X
.trigger.endsweep.action
smu
X
.trigger.endsweep.action =
action
action
The source action at the end of a sweep; set to one of the following values:
0
or
smu
X
.SOURCE_IDLE
: Sets the source level to the programmed (idle) level at
the end of the sweep
1
or
smu
X
.SOURCE_HOLD
: Sets the source level to stay at the level of the last step
X
Source-measure unit (SMU) channel (for example,
smua.trigger.endsweep.action
applies to SMU channel A)
Details
Use this attribute to configure the source action at the end of the sweep. The SMU can be
programmed to return to the idle source level or hold the last value of the sweep.
Example
smua.trigger.endsweep.action = smua.SOURCE_IDLE
Sets SMU channel A to return the source
back to the idle source level at the end of a
sweep.
Also see
(on page 3-32)
smuX.trigger.IDLE_EVENT_ID
This constant contains the idle event number.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
eventID
= smu
X
.trigger.IDLE_EVENT_ID
eventID
The idle event number
X
Source-measure unit (SMU) channel (for example,
smua.trigger.IDLE_EVENT_ID
applies to SMU channel A)
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to
idle events from this SMU.