Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. C / August 2016
7-383
tsplink.master
This attribute reads the node number assigned to the master node. This attribute is not available on the Models
2604B/2614B/2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
masterNodeNumber
= tsplink.master
masterNodeNumber
The node number of the master node
Details
After doing a TSP-Link reset (
tsplink.reset()
), use this attribute to access the node number of
the master in a set of instruments connected over TSP-Link.
Example
LinkMaster = tsplink.master
Store the TSP-Link master
node number in a variable
called LinkMaster.
Also see
(on page 7-385)
tsplink.node
This attribute defines the node number. This attribute is not available on the Models 2604B/2614B/2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Not applicable
Nonvolatile
memory
1
Usage
nodeNumber
= tsplink.node
tsplink.node =
nodeNumber
nodeNumber
The node number of the instrument or enclosure (1 to 64)
Details
This attribute sets the TSP-Link node number and saves the value in nonvolatile memory.
Changes to the node number do not take effect until
tsplink.reset()
from an earlier TSP-Link
instrument is executed on any node in the system.
Each node connected to the TSP-Link system must be assigned a different node number.
Example
tsplink.node = 3
Sets the TSP-Link node for this instrument to
number 3.