Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. C / August 2016
7-253
smuX.trigger.ARMED_EVENT_ID
This constant contains the armed event number.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
eventID
= smu
X
.trigger.ARMED_EVENT_ID
eventID
The armed event number
X
Source-measure unit (SMU) channel (for example,
smua.trigger.ARMED_EVENT_ID
applies to SMU channel A)
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to
armed events from this SMU.
Example
trigger.timer[1].stimulus =
smua.trigger.ARMED_EVENT_ID
Trigger timer when the SMU
passes through the ARM layer.
Also see
(on page 3-32)
smuX.trigger.autoclear
This attribute turns automatic clearing of the event detectors on or off.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Not saved
0 (smuX.DISABLE)
Usage
autoClear
= smu
X
.trigger.autoclear
smu
X
.trigger.autoclear =
autoClear
autoClear
Auto clear setting; set to one of the following values:
0
or
smu
X
.DISABLE
: Turns off automatic clearing of the event detectors
1
or
smu
X
.ENABLE
: Turns on automatic clearing of the event detectors
X
Source-measure unit (SMU) channel (for example,
smua.trigger.autoclear
applies to SMU channel A)