Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-280
2600BS-901-01 Rev. C / August 2016
Bit
Value
Description
B0
Not used
Not applicable.
B1
status.measurement.buffer_available.SMUA
Set bit indicates that there is at least one
reading stored in either or both of the dedicated
reading buffers.
Bit B1 decimal value: 2
Binary value: 0000 0010
B2
status.measurement.buffer_available.SMUB
This bit is only available on Models
2602B/2604B/2612B/2614B/2634B/2636B. Set
bit indicates that there is at least one reading
stored in either or both of the dedicated reading
buffers.
Bit B2 decimal value: 4
Binary value: 0000 0100
B3-B15
Not used
Not applicable.
As an example, to set bit B1 of the measurement event buffer available summary enable register, set
status.measurement.buffer_available.enable =
status.measurement.buffer_available.SMUA
.
In addition to the above constants,
measurementRegister
can be set to the decimal equivalent of
the bit to set. To set more than one bit of the register, set
measurementRegister
to the sum of
their decimal weights. For example, to set bits B1 and B2, set
measurementRegister
to 6 (which is
the sum of 2 + 4).
Example
status.measurement.buffer_available.enable =
status.measurement.buffer_available.SMUA
Sets the SMUA bit of the
measurement event buffer available
summary enable register.
Also see
(on page E-8)
status.measurement.current_limit.*
This attribute contains the measurement event current limit summary registers.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
Models 2601B/2611B/2635B: 2 (All bits
set)
Models
2602B/2604B/2612B/2614B/2634B/
2636B: 6 (All bits set)