Section 11: TSP command reference
Series 3700A System Switch/Multimeter Reference Manual
11-178
3700AS-901-01 Rev. D/June 2018
Example
dmm.func = "twowireohms"
dmm.filter.window = 0.25
dmm.filter.enable = dmm.ON
Set the filter window for 2-wire ohms to 0.25
and enable filtered measurements.
Also see
(on page 11-166)
(on page 11-175)
(on page 11-174)
(on page 11-176)
dmm.fourrtd
This attribute sets the type of 4-wire RTD that is being used.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Reset
DMM reset
Recall setup
Create configuration script
Save setup
0 (dmm.RTD_PT100)
Usage
value
= dmm.fourrtd
dmm.fourrtd =
value
value
The type of four-wire RTD:
▪
dmm.RTD_PT100
or
0
for type PT100
▪
dmm.RTD_D100
or
1
for type D100
▪
dmm.RTD_F100
or
2
for type F100
▪
dmm.RTD_PT385
or
3
for type PT385
▪
dmm.RTD_PT3916
or
4
for type PT3916
▪
dmm.RTD_USER
or
5
for user-specified type
Details
This attribute is only valid when
dmm.func
is set to
"temperature"
and
dmm.transducer
is set
to
dmm.TEMP_FOURRTD
. For all other transducer types, the attribute is set but is not used until the
transducer type is set for four-wire RTD.
All other functions generate an error and return nil when queried. An illegal parameter value error
message is generated if the value specified is not a supported RTD type value as listed in the usage
table.
The four-wire RTD setting is saved with the
dmm.func
function setting, so if you use another
function, then return to
"temperature"
, the four-wire RTD settings you set previously are retained.
Example
dmm.func = "temperature"
dmm.transducer = dmm.TEMP_FOURRTD
dmm.fourrtd = dmm.RTD_PT3916
Sets the type of four-wire RTD for PT3916.