Section 13: Frequently asked questions
Series 3700A System Switch/Multimeter Reference Manual
13-2
3700AS-901-01 Rev. D/June 2018
Figure 145: Channel configuration dialog box
3. If the forbidden box is selected, the channel is forbidden to close. To allow the channel to close,
clear the box.
4. Click
OK
to save the change.
To check the forbidden state of a channel from a remote interface:
You can also clear, check, and set the forbidden state of channels using the following commands:
•
•
(on page 11-62)
•
(on page 11-90)
How do I know if an error has occurred on my instrument?
If you are using TSB Embedded, error messages are displayed in the Instrument Output box when
they occur.
If you are using another remote interface, you might need to use commands to retrieve the error
messages. You can use the commands
(on page 11-235) and
(on page 11-236) to retrieve the number of messages and the text of the messages.
To set the instrument to automatically send generated errors, set
(on page 11-
291) to 1 (enabled).
To set the instrument to automatically send prompts after each command message, set
(on page 11-288) to 1 (enabled).