Section 11: TSP command reference
Series 3700A System Switch/Multimeter Reference Manual
11-310
3700AS-901-01 Rev. D/June 2018
The
width
parameter is valid for digital I/O type channels. Widths of 1, 2, 3, or 4 are supported. If
specified, the scan can read up to four consecutive channels simultaneously, and then saves the
resulting value into the specified reading buffer.
DAC channels are not supported.
Measurement timestamps may vary from channel read timestamps because of the way different
channel types generate reading buffer timestamps.
Example 1
scan.create("3001:3010", "DCV")
For this example, assume
"DCV"
is a
previously defined user configuration for DC
volts. Clears the old scan list and creates a
new scan list with each channel (1 to 10 on
slot 3) using DCV as the overriding DMM
configuration.
Example 2
scan.add("3001:3010", "2wire")
For this example, assume
"2wire" is
a
previously defined user configuration for 2-
wire ohms. Uses
2wire
for all 10 channels
and adds them to the end of the existing
scan list.
Example 3
scan.create("")
Clears the old scan list and creates a new
empty scan list.
Example 4
for chan = 3001, 3010 do
scan.add("" .. chan, "DCV")
scan.add("" .. chan, "2wire")
end
For this example, assume
"DCV"
is a
previously defined user DC volts
configuration, and
"2wire"
is a previously
defined user 2-wire ohms configuration.
Adds channels 3001 through 3010 to the end
of the existing scan list. This loops through
the channels twice, adding channels to the
scan list twice. The first time, it adds
"DCV"
for a channel. The second time, it adds
"2wire"
for that channel. The first
parameter
("" .. chan)
converts the
chan number to a string.
Also see
(on page 11-315)
(on page 11-181)
(on page 11-227)
(on page 5-23)