Series 3700A System Switch/Multimeter Reference Manual
Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018
11-151
An error is generated if:
•
You try to set
dmm.autozero
for
"continuity"
or
"nofunction"
. Error code 1114, "Setting
conflicts with function selected," is generated.
•
You query
dmm.autozero
for
"continuity"
or
"nofunction"
.
nil
is returned with error
code 1114, "Setting conflicts with function selected."
Example
dmm.func = "dcvolts"
dmm.autozero = dmm.ON
dmm.autozero = dmm.AUTOZERO_ONCE
print(dmm.autozero)
timer.reset()
dmm.autozero=2
time=timer.measure.t()
print(time)
Enables autozero for DC volts.
Refreshes the reference points once and sets
autozero to
dmm.OFF
or
0
.
Output:
0.00000
Determines the time delay when autozero is selected.
Also see
(on page 4-3)
(on page 11-166)
(on page 11-179)
(on page 11-209)
(on page 11-219)
(on page 11-308)