
Series 3700A System Switch/Multimeter Reference Manual
Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018
11-331
scan.trigger.channel.stimulus
This attribute determines which trigger events cause the channel actions to occur.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
Instrument reset
Recall setup
Scan reset
Create
configuration script
Save setup
50
(scan.trigger.EVENT_SEQUENCE_COMP
)
Usage
eventID
= scan.trigger.channel.stimulus
scan.trigger.channel.stimulus =
eventID
eventID
Trigger stimulus used for the channel action; see
Details
for possible trigger event IDs
Details
This attribute selects which events cause the channel event detector to enter the detected state. Set
this attribute to
0
to start the channel action immediately at the default setting.
Set
eventID
to one of the existing trigger event IDs shown in the following table.
Trigger event IDs
Trigger event ID
Description
channel.trigger[N].EVENT_ID or 41 to 48
The trigger event generated by the channel trigger
N
.
digio.trigger[N].EVENT_ID or 1 to 14
An edge (either rising, falling, or either based on the
configuration of the line) on the digital input line.
display.trigger.EVENT_ID or 39
The trigger key (TRIG) on the front panel is pressed.
dmm.trigger.EVENT_LIMIT1_HIGH or 53
A DMM trigger event that indicates a measurement has
exceeded the high limit value on limit 1.
dmm.trigger.EVENT_LIMIT1_LOW or 52
A DMM trigger event that indicates a measurement has
exceeded the low limit value on limit 1.
dmm.trigger.EVENT_LIMIT2_HIGH or 55
A DMM trigger event that indicates a measurement has
exceeded the high limit value on limit 2.
dmm.trigger.EVENT_LIMIT2_LOW or 54
A DMM trigger event that indicates a measurement has
exceeded the low limit value on limit 2.
trigger.EVENT_ID or 40
A *trg message on the active command interface. If
GPIB is the active command interface, a GET message
also generates this event.
trigger.blender[N].EVENT_ID or 58 to 59
A combination of events has occurred.
trigger.timer[N].EVENT_ID or 20 to 23
A delay expired.
tsplink.trigger[N].EVENT_ID or 17 to 19
An edge (either rising, falling, or either based on the
configuration of the line) on the TSP-Link trigger line.
lan.trigger[N].EVENT_ID or 29 to 36
A LAN trigger event has occurred.
scan.trigger.EVENT_SCAN_READY or 24
Scan ready event.
scan.trigger.EVENT_SCAN_START or 25
Scan start event.
scan.trigger.EVENT_CHANNEL_READY or 28
Channel ready event.
scan.trigger.EVENT_MEASURE_COMP or 56
Measure complete event.
scan.trigger.EVENT_SEQUENCE_COMP or 50
Sequence complete event.
scan.trigger.EVENT_SCAN_COMP or 26
Scan complete event.
scan.trigger.EVENT_IDLE or 27
Idle event.
schedule.alarm[N].EVENT_ID or 37 to 38
Trigger event generated by the alarm
N
.