Section 11: TSP command reference
Series 3700A System Switch/Multimeter Reference Manual
11-156
3700AS-901-01 Rev. D/June 2018
dmm.calibration.dc()
This function begins a DC adjustment step on the DMM.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
dmm.calibration.dc(
step
)
dmm.calibration.dc(
step
,
value
)
step
The DC adjustment step to perform
value
The value for this adjustment step (if the adjustment step has a value)
Details
This command generates an error if the:
•
Calibration is locked
•
Step is out of sequence
•
Step does not exist
•
Step does not complete successfully
•
Value passed is invalid for the step, out of range, or not needed
Example
For example of use, see
(on page B-23),
(on page B-26),
(on page B-27).
Also see
(on page 11-156)
(on page 11-158)
dmm.calibration.lock()
This function locks calibration to prevent unintended changes.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
dmm.calibration.lock()
Details
Use this command to lock an unlocked calibration. An error is generated if this command is issued
when calibration is already locked.
Once locked, you must unlock calibration before you can perform calibration again.
Save calibration data before locking. Calibration data will be lost if it is not saved before locking.