Series 3700A System Switch/Multimeter Reference Manual
Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018
11-429
Details
Releases a trigger that was asserted with an indefinite pulse width. It also releases a trigger that was
latched in response to receiving a synchronous mode trigger.
Example
tsplink.trigger[3].release()
Releases trigger line 3.
Also see
(on page 11-424)
(on page 11-425)
(on page 11-426)
(on page 11-427)
(on page 11-428)
(on page 11-430)
tsplink.trigger[N].reset()
This function resets some of the TSP-Link trigger attributes to their factory defaults.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
tsplink.trigger[
N
].reset()
N
The trigger line (1 to 3)
Details
The
tsplink.trigger[
N
].reset()
function resets the following attributes to their factory
defaults:
▪
tsplink.trigger[
N
].mode
▪
tsplink.trigger[
N
].stimulus
▪
tsplink.trigger[
N
].pulsewidth
This also clears
tsplink.trigger[
N
].overrun
.
Example
tsplink.trigger[3].reset()
Resets TSP-Link trigger line 3
attributes back to factory default
values.
Also see
(on page 11-426)
(on page 11-427)
(on page 11-428)
(on page 11-430)