Series 3700A System Switch/Multimeter Reference Manual
Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018
11-165
dmm.configure.recall()
This function recalls a user or factory DMM configuration and replaces attributes in the present configuration with
attributes from the recalled version.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
dmm.configure.recall("
configuration
")
configuration
A string that represents the name of the DMM configuration to recall
Details
This command recalls the DMM configuration for one function.
When a configuration is recalled, the function associated with the configuration becomes active.
When you recall a DMM configuration, the existing DMM configuration settings for the function are
replaced by the settings in the recalled configuration. Settings for other functions are not affected. For
example, if the function associated with the configuration was temperature, only temperature settings
are recalled. If a factory configuration is recalled, the function's attributes are set to their factory
default values.
The DMM configuration can be user-defined or factory-defined.
User-defined DMM configurations are set with
dmm.configure.set()
. The factory-defined DMM
configurations are:
▪
"accurrent"
▪
"acvolts"
▪
"commonsideohms"
▪
"continuity"
▪
"dccurrent"
▪
"dcvolts"
▪
"fourwireohms"
▪
"frequency"
▪
"nofunction"
▪
"period"
▪
"temperature"
▪
"twowireohms"