2600AS-901-01 Rev. B / September 2008
Return to
19-37
Series 2600A System SourceMeter® Instruments Reference Manual
Section 19: Remote Commands
digio.writeport
Function
Writes to all digital I/O lines.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage
digio.writeport(data)
data
Value to write to the port; 0 to 16383.
Remarks
• The binary representation of data indicates the output pattern to be written to the I/O port. For
example, a
data
value of 170 has a binary equivalent of 00000010101010. Lines 2, 4, 6 and 8
are set high (1), and the other 10 lines are set low (0).
• Write protected lines will not be changed (see
• The
reset
function does not affect the present states of the digital I/O lines.
• Use the
digio.writebit
and
digio.writeport
commands to control the output state of
the synchronization line while the trigger mode is set to
digio.TRIG_BYPASS
.
Details
See
Also see
,
Example
Sets digital I/O lines 1 through 8 high (binary 00000011111111):
digio.writeport(255)
digio.writeprotect
Attribute
Write protect mask that disables bits from being changed with the
digio.writebit
and
digio.writeport
functions.
Default
0
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
mask = digio.writeprotect
-- Reads write protect mask.
digio.writeprotect = mask
-- Writes write protect mask.
mask
Set to the value that specifies the bit pattern for
write
protect.
Remarks
• Bits that are set to
1
cause the corresponding line to be write protected.
• The binary equivalent of
mask
indicates the mask to be set for the I/O port. For example, a
mask
value of 7 has a binary equivalent 00000000000111. This mask write protects lines 1, 2 and 3.
Details
See
Also see
Example
Write protects lines 1, 2, 3 and 4:
digio.writeprotect
=
15