2600AS-901-01 Rev. B / September 2008
Return to
19-55
Series 2600A System SourceMeter® Instruments Reference Manual
Section 19: Remote Commands
errorqueue.next
Function
Reads an entry from the error/event queue.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage
errorcode, message, severity, node = errorqueue.next()
errorcode
Returns the error code number for the entry.
message
Returns the message that describes the entry.
severity
Returns the severity level (0, 10, 20, 30 or 40).
node
Returns the node number where the error originated.
Remarks
• Entries are stored in a first-in, first-out (FIFO) queue. This function reads the oldest entry and
removes it from the queue.
• Error codes and messages are listed in
.
• If there are no entries in the queue, code 0, “Queue Is Empty” is returned.
• Returned
severity
levels include the following:
•
0 Informational:
Indicates no error: “Queue is Empty”.
•
10 Informational:
Indicates an event or a minor error. Examples: “Reading Available” and
“Reading Overflow”.
•
20 Recoverabl
e: Indicates possible invalid user input. Operation will continue but action
should be taken to correct the error. Examples: “Exponent Too Large” and “Numeric Data Not
Allowed”.
•
30 Serious:
Indicates a serious error and may require technical assistance. Example: “Saved
calibration constants corrupted”.
•
40 Fatal:
Indicates that the Series 2600A is non-operational and will require service. Contact
information for service is provided in
. Examples: “Bad SMU AFPGA image size”,
“SMU is unresponsive”, and “Communication Timeout with DFPGA”.
• In an expanded system, each TSP-Link enabled instrument is assigned a node number.
node
returns the node number where the error originated.
Details
See
(Error and Status Messages) and
(Status Model).
Also see
Example
Reads the oldest entry in the error/event queue:
errorcode, message = errorqueue.next()
print(errorcode, message)
Output:
0.00 Queue Is Empty