2600AS-901-01 Rev. B / September 2008
Return to
19-199
Series 2600A System SourceMeter® Instruments Reference Manual
Section 19: Remote Commands
status.system.condition
status.system.enable
status.system.event
status.system.ntr
status.system.ptr
Attribute
TSP-Link system summary register set.
Default
0
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
Reads condition, enable, event, NTR, and PTR registers:
enablereg = status.system.condition
enablereg = status.system.enable
enablereg = status.system.event
enablereg = status.system.ntr
enablereg = status.system.ptr
Writes to enable, NTR, and PTR registers:
status.system.enable = enablereg
status.system.ntr = enablereg
status.system.ptr = enablereg
Set
enablereg
to one of the following values:
0
Clears all bits.
status.system.EXTENSION_BIT
Sets EXT bit (B0).
status.system.EXT
Sets EXT bit (B0).
status.system.NODE1
Sets node bit 1 (B1).
status.system.NODE2
Sets node bit 2 (B2).
status.system.NODE3
Sets node bit 3 (B3).
status.system.NODE4
Sets node bit 4 (B4).
status.system.NODE5
Sets node bit 5 (B5).
status.system.NODE6
Sets node bit 6 (B6).
status.system.NODE7
Sets node bit 7 (B7).
status.system.NODE8
Sets node bit 8 (B8).
status.system.NODE9
Sets node bit 9 (B9).
status.system.NODE10
Sets node bit 10 (B10).
status.system.NODE11
Sets node bit 11 (B11).
status.system.NODE12
Sets node bit 12 (B12).
status.system.NODE13
Sets node bit 13 (B13).
status.system.NODE14
Sets node bit 14 (B14).
Remarks
• In an expanded system (TSP-Link), These attributes are used to read or write to the system
summary registers.
• Reading a system summary register returns a numeric value whose binary equivalent indicates
which register bits are set. The bits of the system summary register are identified as follows:
• For example, assume the value 9 is returned for the enable register. The binary equivalent is
0000000000001001. This value indicates that bit 0 (EXT) and bit 3 (Node 3) are set.
Details
See
System Summary Event Registers
Also see
status.system2, status.system3, status.system4, status.system5
Example
Sets the extension bit of the system summary enable register:
status.system.enable = status.system.EXT
Bit B0 - EXT bit
Bit B1 - Node 1
Bit B2 - Node 2
Bit B3 - Node 3
Bit B4 - Node 4
Bit B5 - Node 5
Bit B6 - Node 6
Bit B7 - Node 7
Bit B8 - Node 8
Bit B9 - Node 9
Bit B10 - Node 10
Bit B11 - Node 11
Bit B12 - Node 12
Bit B13 - Node 13
Bit B14 - Node 14
Bit B15 - Not used