19-112
Return to
2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands
Series 2600A System SourceMeter® Instruments Reference Manual
smuX.measure.analogfilter
X = SMU channel (a or b)
(Models 2635A and 2636A only)
Attribute
Controls the use of an analog filter when measuring on the lowest current ranges.
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
option = smuX.measure.analogfilter
-- Reads the filter setting.
smuX.measure.analogfilter = option
-- Writes the filter setting.
option
Indicates the filter setting
Where
option
is:
0
filter off
1
filter on
Remarks
• This attribute engages an approximately 1Hz analog filter across the current range elements.
• The analog filter is only active when using the 1nA and 100pA measurement ranges.
Example
Turns off the analog filter:
smua.measure.analogfilter = 0
smuX.measure.autorangeY
X = SMU channel (a or b)
Y = SMU measure function (
v
or
i
)
Where:
v
= voltage,
i
= current
Attribute
Measurement auto range setting.
Default
smuX.AUTORANGE_ON
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
autorange = smuX.measure.autorangei
-- Reads current measure auto
-- range.
smuX.measure.autorangei = autorange
-- Writes current measure auto
-- range.
autorange = smuX.measure.autorangev
-- Reads voltage measure auto
-- range.
smuX.measure.autorangev = autorange
-- Writes voltage measure auto
-- range.
autorange
Indicates whether measurement auto range is active.
Set
autorange
to one of the following values:
0
or
smuX.AUTORANGE_OFF
Disables measurement auto range.
1
or
smuX.AUTORANGE_ON
Enables measurement auto range.
Remarks
• This attribute indicates the measurement auto range state. Its value will be
smuX.AUTORANGE_OFF
when the SMU measure circuit is on a fixed range and
smuX.AUTORANGE_ON
when it is in auto range mode.
• Setting this attribute to
smuX.AUTORANGE_OFF
puts the SMU on a fixed range. The fixed range
used will be the range the SMU measure circuit was currently using.
• Setting this attribute to
smuX.AUTORANGE_ON
puts the SMU measure circuit into auto range
mode. It will remain on its present measure range until the next measurement is requested.
Details
See
Also see
Example
Enables voltage measurement autoranging for SMU A:
smua.measure.autorangev = smua.AUTORANGE_ON