Series 2600A System SourceMeter® Instruments Reference Manual
Section 19: Remote Commands
Example
ConfigPulseIMeasureV(smua, 0, 5, 10, 0.001, 0.080, 1,
smua.nvbuffer1, 1)
Set up a pulse train that will use SourceMeter instrument Channel A. The pulse
amplitude will be 5A and will return to 0A after 1ms. The pulse will remain at 0A
for 80 ms and the voltage limit will be 10V during the pulse. The pulse train will
consist of only 1 pulse and this pulse will be assigned a tag index of 1.
See also
,
Table 19-13
KIPulse TSP test script: ConfigPulseVMeasureI
TSP project name:
KIPulse
TSP test script name:
KIPulse
Function:
ConfigPulseVMeasureI
Usage:
f, msg = ConfigPulseVMeasureI
(smu,bias,level,limit,ton,toff,points,buffer,tag[,sync_in][,sync_out]
[,sync_in_timeout][,sync_in_abort])
Description
Configures a voltage pulse train with a current measurement at each point.
Measurement(s) will be made at the end of the
ton
time. This function does not
cause the specified
smu
to output a pulse train. It simply checks to see if all of the
pulse dimensions are achievable and if so, assigns the indicated
tag
or index to
the pulse train.
The InitiatePulseTest(tag)
and
InitiatePulseTestDual(tag)
functions are used to initiate a pulse train
assigned to a valid tag.
Table 19-12 (cont.)
KIPulse TSP test script: ConfigPulseIMeasureV
TSP project name:
KIPulse
TSP test script name:
KIPulse
Function:
ConfigPulseIMeasureV
Usage:
f, msg = ConfigPulseIMeasureV
(smu,bias,level,limit,ton,toff,points,buffer,tag[,sync_in][,sync_out]
[,sync_in_timeout][,sync_in_abort])