2600AS-901-01 Rev. B / September 2008
Return to
19-165
Series 2600A System SourceMeter® Instruments Reference Manual
Section 19: Remote Commands
Attribute
Operation status instrument summary register set.
Default
0
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
Reads condition, enable, event, NTR, and PTR registers:
operreg = status.operation.instrument.condition
operreg = status.operation.instrument.enable
operreg = status.operation.instrument.event
operreg = status.operation.instrument.ntr
operreg = status.operation.instrument.ptr
Writes to enable, NTR, and PTR registers:
status.operation.instrument.enable = operreg
status.operation.instrument.ntr = operreg
status.operation.instrument.ptr = operreg
Set
operreg
to one of the following values:
0
Clears all bits.
status.operation.instrument.SMUA
Sets SMUA bit (B1).
status.operation.instrument.SMUB
Sets SMUB bit (B2).
status.operation.instrument.TRIGGER_BLENDER
Sets TRIGBLND bit (B10).
status.operation.instrument.TRGBLND
Sets TRGBLND bit (B10).
status.operation.instrument.TRIGGER_TIMER
Sets TRGTMR bit (B11).
status.operation.instrument.TRGTMR
Sets TRGTMR bit (B11).
status.operation.instrument.DIGITAL_IO
Sets DIGIO bit (B12).
status.operation.instrument.DIGIO
Sets DIGIO bit (B12).
status.operation.instrument.TSPLINK
Sets TSPLINK bit (B13).
status.operation.instrument.LAN
Sets LAN bit (B14).
operreg
can also be set to the decimal weight of the bit to be set. Examples:
To set bit B1 (SMUA), set
operreg
to 2 (2
1
).
To set bit B2 (SMUB), set
operreg
to 4 (2
2
).
To set bit B12 (DIGIO), set
operreg
to 4096 (2
12
).
To set two bits, set
operreg
to the sum of the decimal weights of both bits:
To set bits B1 and B2, set
operreg
to 6 (2 + 4).