2600AS-901-01 Rev. B / September 2008
Return to
19-85
Series 2600A System SourceMeter® Instruments Reference Manual
Section 19: Remote Commands
localnode
Use the attributes and functions in this section to set the power line frequency, control prompts (on
and off), control error messages, (show
and hide), to access global variables, and to run test
scripts.
There are two different ways to use the
localnode
object.
•
Send commands from the local node.
•
Send commands from a remote node.
To send commands from the local node, use the
localnode
element:
localnode.linefreq = 50
NOTE Use the node reference to send commands from a remote node, do
not use the
localnode
element.
node[n].linefreq = 50
NOTE Some
localnode
commands cannot be run on the local node. They
must be run from a remote node.
localnode.autolinefreq
Attribute
Automatic power line frequency detection control.
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
flag = localnode.autolinefreq--
Read auto line frequency detection
-- setting.
localnode.autolinefreq = flag
-- Set auto line frequency detection
-- mode.
flag
The auto line frequency detection setting.
Set
flag
to one of the following values:
true
Enable automatic line frequency detection at start-up.
false
Disable automatic line frequency detection at start-up.
Remarks
• When this attribute is set to
true
, the power line frequency is detected automatically the next
time the Series 2600A powers up.
• After the power line frequency is automatically detected at power-up, the
localnode.linefreq
attribute will be set automatically to 50 or 60.
• If the
localnode.linefreq
attribute is explicitly set,
localnode.autolinefreq
will be automatically set to
false
.
Also see
Example
Disable automatic power line frequency detection:
localnode.autolinefreq = false