19-152
Return to
2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands
Series 2600A System SourceMeter® Instruments Reference Manual
status.measurement.condition
status.measurement.enable
status.measurement.event
status.measurement.ntr
status.measurement.ptr
Attribute
Measurement event register set.
Default
0
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
Reads condition, enable, event, NTR, and PTR registers:
measreg = status.measurement.condition
measreg = status.measurement.enable
measreg = status.measurement.event
measreg = status.measurement.ntr
measreg = status.measurement.ptr
Writes to enable, NTR, and PTR registers:
status.measurement.enable = measreg
status.measurement.ntr = measreg
status.measurement.ptr = measreg
Set
measreg
to one of the following values:
0
Clears all bits.
status.measurement.VOLTAGE_LIMIT
Sets VLMT bit (B0).
status.measurement.VLMT
Sets VLMT bit (B0).
status.measurement.CURRENT_LIMIT
Sets ILMT bit (B1).
status.measurement.ILMT
Sets ILMT bit (B1).
status.measurement.READING_OVERFLOW
Sets ROF bit (B7).
status.measurement.ROF
Sets ROF bit (B7).
status.measurement.BUFFER_AVAILABLE
Sets BAV bit (B8).
status.measurement.BAV
Sets BAV bit (B8).
status.measurement.OUTPUT_ENABLE
Sets OE bit (B11).
status.measurement.OE
Sets OE bit (B11).
status.measurement.INSTRUMENT_SUMMARY
Sets INST bit (B13).
status.measurement.INST
Sets INST bit (B13).
measreg
can also be set to the decimal weight of the bit to be set. Examples:
To set bit B0 (VLMT), set
measreg
to 1 (2
0
).
To set bit B1 (ILMT), set
measreg
to 2 (2
1
).
To set bit B8 (BAV), set
measreg
to 256 (2
8
).
To set more than one bit of the register, set
measreg
to the sum of their decimal weights. For
example, to set bits B0 and B8, set
measreg
to 257 (1 + 256).