Series 2600A System SourceMeter® Instruments Reference Manual
Section 4: Source-Measure Concepts
CAUTION
This maximum duty cycle equation is an approximation. In general, if the
duty cycle calculation yields a number > 90%, then DC under those
conditions should not cause the SourceMeter instrument to overheat.
However, if the calculation yields a number < 10%, the calculated duty cycle
should not be exceeded by more than 0.5% to avoid potential overheating.
Examples
Example 1:
Using a Model 2611A to charge a 5V battery with 1.5A, while operating at 50°C ambient
temperature; what is the maximum duty cycle?
Assuming the 20V range will be used to measure the voltage:
DC
MAX
≤ 47.0%
Example 2:
Using a Model 2602A to pulse 10A of current from a bias level of 500mA, into a very low
impedance (100 m
Ω
), while operating at 40°C ambient temperature; what is the maximum duty
cycle?
Assuming the 1V range will be used to measure the voltage:
DC
MAX
≤ 4.7%
Table 4-2
Model 2601A/2602A Maximum Duty Cycle equation constants
Constant
100mV range
1V range
6V range
40V range
P
CS
56
56
56
56
V
OA
18
18
18
55
Table 4-3
Model 2611A/2612A/2635A/2636A Maximum Duty Cycle equation constants
Constant
200mV range
2V range
20V range
200V range
P
CS
56
56
56
56
V
OA
40
40
40
220
DC
MAX
56
20
–
(
)
40
5
( )
–
(
)
0
( )
–
40
5
( )
–
(
)
1.5
(
)
-------------------------------------------------------------------
100
×
2
≤
DC
MAX
56
10
–
(
)
18
0.1
(
)
0.5
(
)
–
(
)
0.5
(
)
–
18
10
(
)
0.1
(
)
–
(
)
10
(
)
-------------------------------------------------------------------------------------------
100
×
2
≤