19-36
Return to
2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands
Series 2600A System SourceMeter® Instruments Reference Manual
digio.trigger[N].wait
Replace N with the number of the digital I/O trigger line: 1 to 14.
Function
Waits for a trigger.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage
triggered = digio.trigger[n].wait(timeout)
timeout
Specifies
the
time-out value in seconds.
triggered
A customized variable that stores the value
true
if a trigger is detected or
false
if a trigger is not
detected during the time-out period.
n
The number of the trigger line.
Remarks
• This function waits up to the
timeout
value in seconds for an input trigger. If one or more trigger
events are detected since the last time
digio.trigger[n].wait
or
digio.trigger[n].clear
was called, this function immediately returns. After waiting for a
trigger with this function, the event detector is automatically reset and rearmed. This is true
regardless of the number of events detected.
Details
See
Also see
Example
Waits up to three seconds for a trigger to be detected on trigger line 4, then displays if the trigger was
detected:
triggered = digio.trigger[4].wait(3)
print(triggered)
Output:
false
A trigger was not detected.
true
A trigger was detected.
digio.writebit
Function
Sets a digital I/O line high or low.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage
digio.writebit(n, data)
n
The digital I/O line number (1 to 14).
data
The value to write to the bit; 0 (low) or 1 (high).
Remarks
• If the output line is write protected, via the
attribute, the command will
be ignored.
• The
reset
function does not affect the present state of the digital I/O lines.
• Use the
digio.writebit
and
digio.writeport
commands to control the output state of
the synchronization line when the trigger mode is set to
digio.TRIG_BYPASS
.
Details
See
Also see
,
Example
Sets digital I/O line 4 low (0):
digio.writebit(4, 0)