Pressure Loss
According to OIML R49 the maximum pressure loss must not exceed 0.63 bar in range Q
1
up to and incl. Q
3
and max 1.0 bar at Q
4
respectively. The pressure loss is without anti-pollution check valve.
The pressure loss in a sensor increases with the square of the flow and can be stated as:
p
kv
Q
where:
Q
= volume flow rate [m³/h]
kv
= volume flow rate at 1 bar pressure loss [m³/h]
p
= pressure loss [bar]
Graph
Q
3
Nom. diameter
kv
Q @ 0.63 bar
[m³/h]
[mm]
[m³/h]
A
1.6
DN15 & DN20
3.2
2.5
B
2.5 & 4 & 6.3
DN20 & DN25
13.4
10.6
C
10 & 16
DN40 & DN50
40
32
D
25
DN65
102
81
E
40
DN80
179
142
Table 6 Pressure loss table
Diagram 2 Pressure loss chart
0,01
0,1
1
0,1
1
10
100
1000
p
[
b
ar
]
Flow [m³/h]
p ULTRAFLOW
®
24
A
B
C
D
E