TECHNICAL DESCRIPTION
MULTICAL® 62
Ka st up A/“ •
Technical
Des iptio •
5512-1036-GB_D1
•
.2016
84
13
Verification
13.1
High resolution volume for test
If high resolution volume reading (V1HighRes) is needed during test or verification of the meter, it can be
initialized as follows:
-
Lift the calculator top off the connection base and wait until the display goes blank.
-
Activate the sub-key
and place the calculator top on the connection base again, press the key
for approx. 8 seconds, until the display becomes active in HighRes mode.
-
The display remains active in HighRes verification mode, until a key is activated or the calculator
top is reset.
Example:
V1
V1HighRes
0.001 m3
0.0001 L
0.01 m3
0.001 L
0.1 m3
0.01 L
1 m3
0.1 L
Example of a high resolution volume (V1HighRes) reading:
In the example below the display start value is 573.24 m
3
(v1). When HighRes mode becomes active the
display changes to high resolution, and the litre reading appears. Now a pulse value for verification can be
entered;
here with a value of
20.205.
00573.24 m
3
(0057) 3.240000 m
3
3240.000 L
+ 20.205 L
3260.205 L
Please note: V1HighRes is updated periodically every 10 seconds.
13.1.1
Data reading of high resolution volume
The register ‘HighRes’
can be data read using ID = 239.
By data reading, the measuring unit and value will appear correctly, irrespective of meter size.