Mirus for H2 Classic GrainGage
67
H2 Classic GrainGage Calibration
For example, Sample One has the following moisture measurements:
•
Benchtop sensor: 18%
•
GrainGage: 16.8%
The offset would be 18% − 16.8% = +1.2%
12. Average the sample offsets.
For example, if your benchtop sensor says Sample One is +1.2% more than the GrainGage measurement
and Sample Two is +0.8%, the average offset is +1.0%.
13. Add the average offset to each moisture point with exception of the 0 and 1 moisture points (never
adjust 0 or 1). Subtract for a negative average offset. Do not adjust voltages.
In the example above, readings from the benchtop sensor indicate that the samples have, on average, 1%
higher moisture than the percentages in the previously calculated moisture curve.
The following tables illustrate this type of adjustment:
Original Curve
Adjusted Curve
Moisture
Voltage
Moisture
Voltage
0.0
0.000
0.0
0.000
1.0
0.400
1.0
0.400
8.0
0.793
9.0
0.793
11.0
0.892
12.0
0.892
14.0
0.990
15.0
0.990
17.0
1.089
18.0
1.089
20.0
1.187
21.0
1.187
36.0
1.713
37.0
1.713
40.0
1.845
41.0
1.845
Add or subtract
the offsets here.
As needed, add or
remove points in the
curve here.
Summary of Contents for HarvestMaster mirus H2
Page 1: ......
Page 7: ...CHAP TER ONE Mirus Ins tallation...
Page 11: ...Mirus and GrainGage S etup CHAP TER T WO...
Page 32: ...H2 Clas sic GrainGage Diagnos tic s and Aler t s CHAP TER THREE...
Page 42: ...GrainGage Calibration CHAP TER FOUR...
Page 71: ...Field Maps CHAP TER FIVE...
Page 98: ...Har ves t Mode CHAP TER SIX...
Page 120: ...Standard Grain Information APPENDIX A...
Page 123: ...Troubleshooting APPENDIX B...
Page 129: ...Limited Warrant y APPENDIX C...