56
H2 Classic GrainGage Calibration
The following steps explain how to begin a chamber calibration by using moisture and test weight.
15. Pour the grain sample into the GrainGage hopper to
simulate harvest.
16. Enter the Known Moisture and Test Weight.
17. Tap
Click to run a sample
.
Note: After the first sample is cycled the values will not
change.
The GrainGage cycles the grain, and Mirus displays the
voltage, moisture percentage,
and
test weight.
18. Collect the same sample and pour it into the
GrainGage again.
19. Repeat the calibration cycle with the same test weight
(steps 15–19) at least two more times for a total of three
times with this grain sample.
If the moisture content of your samples does not cover a
range of at least 3%, Mirus displays an Insufficient Moisture
Range warning.
•
Tap
Ignore
to continue with the current samples.
•
Tap
Return
to add more samples with a variety of
moisture percentages.