4
handpiece. In this case, connect the plug at the end of
the integrated cable directly to the generator.
CAUTION:
If the VAPR generator display does not match
the default settings specified below, consult the user
manual for further instructions.
ELECTRODE
GEOMETRY
3.5mm 90°
Suction Electrode
LDS Electrode
VAPR SYSTEM
VAPR;
VAPR II
VAPR
3
VAPR;
VAPR
II
VAPR
3
Default mode
V2
V3
BV2
V3
Default
Vaporization
Power (W)
160
160
160
240
Default
Desiccation
Power (W)
120
120
120
120
Maximum
Vaporization
Power (W)
200
200
200
260
Maximum
Desiccation
Power (W)
160
160
160
160
Minimum
Cannula Size
5.5mm ID
cannula
6.3mm ID or
DePuy Mitek
6mm cannula
P/N# 214660,
214661
ELECTRODE
GEOMETRY
LPS Electrode/
VAPR S
90
Electrode/
Premiere90 Electrode
VAPR SYSTEM
VAPR;
VAPR II
VAPR 3
Default mode
BV2
V3
Default
Vaporization
Power (W)
160
240
Default
Desiccation
Power (W)
120
120
Maximum
Vaporization
Power (W)
200
260
Maximum
Desiccation
Power (W)
160
160
Minimum
Cannula Size
5.5mm ID cannula
ELECTRODE
GEOMETRY
S
50
Electrode
VAPR SYSTEM
VAPR;
VAPR II
VAPR 3
Default mode
V2
V2
Default
Vaporization
Power (W)
60
60
Default
Desiccation
Power (W)
45
45
Maximum
Vaporization
Power (W)
90
90
Maximum
Desiccation
Power (W)
60
60
Minimum
Cannula Size
Not Applicable,