Chapter 1. Installation Overview
STERRAD
®
100S Installation Guide
8
Environmental Conditions
Transport and Storage
Ambient Temperature
-40 to +70° C (50 to 104° F)
Relative Humidity
10 to 100% including condensation
Atmospheric Pressure
500 hPa to 1060 hPa
Operation
Ambient Temperature
+10 to 40° C (50 to 104° F)
Altitude
-100 to 3000 m above msl (-330 to 10,000 ft above msl)
Relative Humidity
0 to 95% RH, non-condensing
Atmospheric Pressure
700 hPa to 1060 hPa
Surface
The floor at the installation is flat and level to + 5 mm/M.
Installation Space Requirements
In operation, the system shall not be placed closer than 50
mm (2 in) to a wall at the rear and/or sides of the system.
The system should be installed in a space of sufficient size
to permit access to all four sides of the system when the
system is rolled away from the rear and/or side walls a
distance of 3 meters (10 ft), the extent of the power cord.
Service access should allow for a minimum clearance of
one meter (3 ft) on all sides of the system.
Mobility
The sterilizer is designed to be free standing and self
contained with no required connections other than the
electrical power cord. The sterilizer is equipped with
wheels that allow it to be rolled easily from one location to
another. The unit is also equipped with adjustable feet at
the front of the unit if leveling is required.
Summary of Contents for ASP STERRAD 100S
Page 1: ...STERRAD 100S Sterilizer Installation Guide ...
Page 5: ...STERRAD 100S Installation Guide 5 Chapter 1 Installation Overview ...
Page 14: ......
Page 15: ...STERRAD 100S Installation Guide 15 Chapter 2 For Your Safety ...
Page 20: ......
Page 21: ...STERRAD 100S Installation Guide 21 Chapter 3 Installation ...
Page 23: ...Chapter 3 Installation STERRAD 100S Installation Guide 23 Figure 1 Installation Diagram ...