Pre-Installation Procedure-To Be Done by the Customer
The customer must identify the location for the sterilizer and perform
steps 1-4 below prior to any other installation procedures.
Figure 3. Optional Mounting Plan
Note: There is a drilling template included in the kit.
It has the proper dimensions and drilling
positions.
1. Position the rear bracket on the floor about 2 inches away from the
wall (1-1/16 inches minimum) and mark the location of the holes for
drilling. If there is an outlet in the wall at the desired installation
location, add at least 3 inches to the distance.