Chapter 3. Installation
STERRAD
®
100S Installation Guide
38
Install Printer Ribbon and Paper
1. Install the printer ribbon.
2. Install a roll of paper in the printer. Place the rest of the paper in the
accessory box for shipment to the customer.
Power Connection
1. Verify that the main circuit breaker located on the AC interface
assembly is in the OFF position.
2. Plug the vacuum pump AC power connector into the designated
receptacle on the AC enclosure.
3. Check all circuit breakers located on the AC enclosure. The circuit
breakers are on the side of the unit and accessible with the panels in
place. Verify that all circuit breakers are ON. Movement during
shipping may cause the circuit breakers to trip.
Figure 8. Circuit Breakers.
4. Check that all of the connectors throughout the entire machine are in
place and correctly mated. Some of the connectors may have worked
loose in shipment.
5. Remove both control enclosure panels and ensure that all cables
attached to the connector distribution PCBs are secure. Verify that
the 3 PCBs are fully inserted into their sockets.
6. Plug the system into the outlet.