Chapter 3. Installation
STERRAD
®
100S Installation Guide
35
6. Connect orange wire #90 to CR3-NC 96.
7. Connect black wire #29 to CR3-T1.
8. Connect blue wire #30 to CR3-T2.
9. Connect red wire #31 to CR3-T3.
10. Install CR3 to CR2 by hooking CR3 into the cover of CR2 and
inserting pins into CR2-T1, T2, and T3. Secure the terminal screws.
11. Attach the configuration PWA to the standoffs in the control
enclosure and plug the connectors into the PWA.
Power Plug
1. If not already installed, install a plug to the power cord for the
correct voltage for your country. Connect as shown in the table
below.
Wire Number
Color and Number
P80-Connect
1 Black
1
P80-X
2 Black
2
P80-Y
3 Black
3
P80-Z
4 Black
4
P80-W
5 Green/Yellow
P80-Ground
Control Panel Bezel
1. Remove the plastic cover from the INSIDE of the clear plastic
window. DO NOT remove the outer protective plastic overlay.
2. Press the control bezel into place on the front of the control panel
until pins lock in the detents and the panel is flush with the front
surface of the bezel.