Chapter 2. For Your Safety
STERRAD
®
100S Installation Guide
19
Device Safety
•
The chapter on Preparing Items To Be Sterilized in the STERRAD
100S User's Guide contains information about materials and devices
that can be processed by the sterilizer. Make sure your customer is
familiar with that section prior to using the sterilizer.
•
Metal objects must not come into contact with the chamber walls, the
door, or the electrode. Contact with the walls, door, or electrode
could damage the sterilizer or the instruments.
•
Warn the customer that they must not change the power source
without checking the electrical phase rotation. Also, prior to
relocating the sterilizer to a new power source, electrical phase
rotation should be checked. Failure to verify and match phase
rotation may cause damage to the sterilizer.