8.4.2 Replacing the Source Lamp
When the filament of the light-source lamp has burned out, replace it, following the
procedure below. Please ask our service personnel concerning information about how to
procure the replacement lamps.
1.
Turn off the power to the instrument.
Wait for 30 minutes or longer after the lamp is turned off. Then, go on to the next
step.
To avoid burns, allow the lamp to cool down for at least 30 minutes
before removing it. Never touch the lamp with bare hands because
it is very hot immediately after the filament burns out.
WARNING
2.
Remove the left-side panel of the main unit using a hexagonal wrench.
Be careful not to put your hand under the base plate. You might get
your fingers caught.
WARNING
3.
Turn the lock of the lamp-exchange door to OPEN using a coin.
4.
Remove the lamp exchange door.
8-8
TMPM5200-2