6
ANALYSIS OPERATION
6.11.1 Display Parameters
The display parameters for CITS images are specified in this window. CITS images are
displayed in a size of 128
×
128 pixels.
The parameters to be set are as follows.
Display Mode
Display modes are selected with these radio buttons.
•
Normal
Current images are displayed directly.
•
Difference
Images of the current difference between neighboring frames are displayed.
•
Normalized Difference
Images of the normalized current difference between neighboring frames are
displayed.
•
Normalized d (ADC)
Images of the normalized current difference between neighboring frames based on
the original AD values are displayed.
•
Conductance
Derivative (dI/dV) images are displayed.
Show Bias Voltage
This function enables the bias voltage used to acquire an electric current image to be
either displayed or not displayed.
You can select one from the following two modes:
Overlay of displayed text (The area around text is not erased.) and Opaque (area around
text is erased) Off (not displayed)
Number to display:
You can select the number of images to be displayed (up to 16 images).
TMPM5200-2