Range setting of Vbe / Vce / Ic when VIEW MODE=WAVE
When VIEW MODE
=
WAVE, range settings for Vbe / Vce / Ic of each waveform depend on the
collector mode as shown below:
In addition, there are limitations below for range setting when setting HORIZONTAL source and
VERTICAL source.
To the way in a right figure, the range being set is displayed in the area under the menu.
(When VIEW MODE is WAVE)
COLLECTOR
SUPPLY MODE
Range
Max. value
Vbe All MODEs
5.00V, 2.00V, 1.00V, 500mV, 200mV, 100mV, 50.0mV
5.00V
Vce High Voltage
500V, 200V, 100V, 50.0V, 20.0V, 10.0V, 5.00V, 2.00V, 1.00V,
500mV, 200mV, 100mV, 50.0mV
500V
High Current
5.00V, 2.00V, 1.00V, 500mV, 200mV, 100mV, 50.0mV
50.0V
Ic
Leakage
2.00mA, 1.00mA, 500uA, 200uA, 100uA, 50.0uA, 5.00uA, 2.00uA,
1.00uA, 500nA, 200nA, 100nA, 50.0nA, 20.0nA, 10.0nA, 5.00nA,
2.00nA, 1.00nA
2.00mA
High Current
200A, 100A, 50.0A, 20.0A, 10.0A, 5.00A, 2.00A, 1.00A, 500mA,
200mA, 100mA
* The setting range is different according to the setting range of MAX
PEAK POWER WATTS.
200A
Others
High Voltage
2.00A, 1.00A, 500mA, 200mA, 100mA, 50.0mA, 20.0mA, 10.0mA,
5.00mA, 2.00mA, 1.00mA, 500uA, 200uA, 100uA, 50.0uA, 20.0uA,
10.0uA, 5.00uA, 2.00uA, 1.00uA
2.00A
HORIZONTAL source setting
STEP
BASE
-
EMITTER
COLLECTOR
Vbe Max. value fixed
Vbe=5.00 V/div
Vce=500V/div
Range setting with FUNCTION knob and
range setting with VERTICAL knob
operate together; i.e., the same range is
generated.
FUNCTION knob allows independent
setting.
Vce
FUNCTION knob allows independent
setting.
Range setting with FUNCTION knob and
range setting with VERTICAL knob
operate together; i.e., the same range is
generated.
VERTICAL source setting
STEP
COLLECTOR or EMITTER
Ic
Max. value
fixed
2.00 A/div
Range setting with FUNCTION knob and range setting with VERTICAL knob operate
together; i.e., the same range is generated.