Capter3
Function
Setting and Basic Operation
Curve Tracer CS-5000 Series Instruction Manual
3-71
< The first page of MEASUREMENT SETUP submenu >
NO.
Submenu
Function description
F1
VARIABLE at RECALL
Sets whether Variable setting of the collector supply output is
restored when recalling the setup.
- RECALL: The collector supply output is restored to the stored
Variable setting value.
- 0.0
%
:
The collector supply output is set to 0.0
%
regardless of
the stored Variable setting value.
F2
VARIABLE at INTERLOCK
Sets whether Variable setting of the collector supply output is held
when interlock occurs.
- HOLD: When interlock occurs, the collector supply output stops
but Variable setting value is held.
When interlock is released, the collector supply output is
restarted in accordance with Variable setting value.
- 0.0
%
:
When interlock occurs, the collector supply output is
reset to 0.0
%
.
When interlock is released, the collector supply output is
held to 0.0
%
.
F3
COLLECTOR SUPPLY OUTPUT
at MEASURE STOP
Sets the status of the collector supply output in measurement
STOP status. It is usually recommended to use like LEAK ONLY of
the default configuration among three following kinds of settings for
safety.
- OFF:
In all modes (AC,RECTIFIED SINE,DC,LEAKAGE), stop
the output of the collector supply.
It enters the state of Over load by the rush electric current
when it changes to might the stop of the measurement, the
influence of the change of the collector voltage of beginning
on the measurement result for LEAKAG E mode, and the
state of output ON, there is a possibility that the
measurement stops, and use it after confirming there is no
problem.
- LEAK ONLY (default setting):
Stop the output of the collector supply. However, do not
stop the collector supply output while the measurement
even is stopping because the change of the output of the
collector supply influences the measurement only at
LEAKAG E mode
- ALL:
Continue the output of the collector supply.
Set only when the change of the collector supply output
according to the measured change influences the
measurement.
F4 to F5
(Not used)
F6
NEXT PAGE
The menu is switched to NEXT PAGE by pushing F6 key.