Capter4
Characteristics
Measurement
Curve Tracer CS-5000 Series Instruction Manual
4-37
(13) Press CONFIGURATION key.
In HIGH CURRENT, the setting is fixed to Collector/StepGen/Common
Checking pulse waveform in WAVE mode
(14) Press VIEW/PULSE key.
(15) Press F1 key to set VIEW MODE to WAVE.
WAVE 1st
(16) Press F2 key to set to Vce.
(17) Rotate FUNCTION knob to set Vce range. Make proper setting in accordance with device
measurement value.
WAVE 2nd
(18) Press FUNCTION knob to set to Ic.
(19) Rotate FUNCTION knob to set Ic range. Make proper setting in accordance with device
measurement value.
(20)Set PULSE WIDTH to
400μs
. Make proper setting depending on the device.
(21) Set MEASUREMENT POINT to
300μs
. Make proper setting depending on the device.
(22) Move the cursor to the measurement point where waveforms of WAVE 1st and WAVE 2nd
are stable.
Starting measurement
(23) Set HIGH CURRENT breaker to ENABLED.
(24) Gradually rotate VARIABLE and apply the current to the device.
After checking WAVE waveform
(25) Press F1 key and set VIEW MODE to TRACE.